A little very basic radiation information for
those who care and don't already know.....
DOSIMETERS are one of the most usefull of radiation instruments. Radiation is
undetectable by humans as it has no smell, feel or other indicator of it's
presence. It causes cumulative damage to the body over time. The amount you
receive in a period of time is known as DOSE. If you are exposed to a RATE of 10
rads for 2 hours you then receive a DOSE of 20 rads. To measure RATE one needs a
RATE meter, such as a Geiger Counter or Survey Meter. More on those in another
thread.
To measure your personal DOSE requires a DOSIMETER (dose meter). There are
several types but the one of interest to us here is a "pen" or
electrostatic type. They have been made since the forties and remain essentially
the same today, still being manufactured. They can be purchased new for about
$100 each or surplus for just a few dollars. They are very rugged and rarely
fail. Most exhibit a little leakage but it is easily compensated for. Leakage is
a manufacturing defect from what I have read. Those few with problems are easily
identified and can either be used with the known fault or trashed. I leak test
all of mine and mark the leak rate on the instrument for future reference and
compensation. Those with high leak rate get tossed while the low leaakage ones
are good to use.
The principle of these is simple. An electrostatic charge is impressed on the
instrument until the indicator rests at zero. If radiation penetrates the unit,
a small amount of that charge is lost in proportion to the amount of radiation.
As charge is lost, the indicator moves up scale. Leakage, the most common fault,
is when the indicator moves slowly up scale when no radiation is present (other
than background). To test for leakage, charge each instrument to a zero reading
and place them in a location where they will not be disturbed. Read them every
12 hours and record where the indicator is. After several 12 hour periods (or
other convienent tiem) dust off the calcutator and figure the rate of change per
12 hours. Use felt tip marker and put the leak rate on each instrument. Just
that simple!!! Example: After 36 hours, a reading of 12 rads is shown. Leak
rate, therefore is 36/12=3 periods. 12/3=4 rads per period Mark as 4/12. In use,
the leak rate is subtracted from the reading during exposure. Leak and exposure
both need to be for same time period. A few refuse to charge, have
"sticky" indicator, cloudy or internally dirty viewing window; these
get tossed.
Dosimeters are available in several ranges. I have Civil Defense ones that are
0-200 RADS (Gold Clip, CD V-742), 0-20 RADS (Green Clip CD V-730 ) and 0-200
MILLIRADS (Black Clip CD V-138 ). The instruments are most accurate at 1/2
scale. If you accumulating a DOSE of 100 RADS in a day or two, the Shit has
truly splattered all over the fan!! These 0-200 R are the range most commonly
sold and the cheapest. By all means, get a few, test and mark them for leakage
and put in your stash. The lower range ones are likely to be more useful in
anything short of terrorist attack close by so try to get a few of those as
well. Always carry the higher range AND lower range when exposure is possible so
that if you exceed your low range one you can have some idea of by how much!
There are nifty electronic dosimeters out there. They are more expensive,
require batteries and are more complex, therefore, maybe, less reliable. The CD
type should be your first purchase and ALWAYS kept around as a reliable back-up.
CHARGERS:
Many variations exist. CD V-750 is the yellow box you often see for sale. They
use 1 "D" cell and also provide illumination. They are a simple and
reliable electronic circuit that dates to the early sixties. Light is provided
by a one cell flash lite bulb. Most common problem is leaking batteries. DO NOT
store with battery installed!! If you buy one that has a lot of corrosion, set
aside for parts and buy another.
Lamp failure: Merely requires a new bulb.
"Noisy" pot. The adjusting pot can (and often does) get scratchy, just
like old volume controls did. A small squirt of electronic control cleaner will
often cure it.
Circuit failure: Set aside for parts. The transistor used hasn't been made for
30 years! Substitutions are possible if you are into electronic repair and have
the time. This sort of failutre is actually very rare.
Self powered type PP1578A/PD as used by the Army. I recently obtained a number
of these and all check good. Last Army "calibration" was 1984. No
battery required, simple and rugged. Disadvantage is they do not provide
illumination. A window or flashlight is required. Due to not needing a battery
they are great for SHTF!! We all know how difficult it could get to obtain
batteries if the balloon ever goes up!!
Use of the above goodies:
An excellent reference book is the FEMA "Radiation Safety In Shelters"
www.radmeters4u.com/FEMA_book/cover.htm
circa 1983. They can also be purchased for around $15. which I STRONGLY
suggest!! (Any idea how long the internet will operate after a SHTF event?)
More later!!
Diplomacy is the art of saying "nice
doggie" while you're looking for a rock." Will Rogers
Excellent info. Thanks
Edit to add. http://www.hosfelt.com has kit
with 2 dosimeters and charger for 14.95 also a couple of different surplus
radiation detectors
Dosimeters and other radiation response supplies
here.
Ki4u
melbo
I wear one 12 hours a shift, and the gov't says that
I've been clean for 14.5 years. Yeah right
M
I'm going to spring for the book. I know amazingly
little, other than hearsay, about radiation. While radiation talent is on line,
let me ask a barely related question. I recently acquired some huge slabs(150
lb. each) of very high quality(pure) lead for bullet casting. Only after closing
deal, seller advised, it was shielding from large x-ray unit. Does lead repell
x-rays or absorb it? Is it possible that melting it would free 50 years worth of
x-rays? Is there a way( or need )of testing it?
Your lead will be just fine to use.
X-rays are like Gamma - an electromagnetic radiation that can not be stored. The
lead does not become radioactive nor is it "contaminated".
more to come |